It was reported that the customer opened the package of the ureteral balloon dilation catheter.And found that it could not be pressurized after normal use, and then took out the balloon for testing and found that the tip was leaking.The user experienced prolonged operation time and has been exposed to harmful substances, blood, or bodily fluids.It was unknown what medical intervention was provided.
|
|
The reported event is confirmed cause unknown.Visual evaluation noted the sample was received in a closed ziploc bag identified with a biohazard label and sterilized by eo label; the bag was handwritten the customer complaint number.Inside of the bag there was the pouch with the unit label that reads ref: 998806.Lot: bmhnfm26.Inside of the pouch it was polyethylene bag with the balloon catheter with the stop cock connected to the balloon hub, the refold tool was located in the proximal of the balloon and there was no guard.The balloon hub of the catheter reads "8mm x 6cm" and "30 atm." visual inspection was performed to the balloon catheter and it was identified a crack in the balloon hub connector that was attached to the stop cock.The stopcock was removed and it was noticed that the crack started from the top of the luer up to the hub wings.Distal tip was visually inspected, and no abnormalities were observed.The exact root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be poor inflation lumen design (lumen id undersized, inflation lumen wall thickness undersized).However, there was insufficient information to confirm this potential root cause.A dhr review did not show any problems or conditions that would have contributed to the reported event.The instructions for use were found adequate and state the following: ¿x-force® balloon dilation catheter instructions for use description the x-force® balloon dilation catheter is a dual lumen catheter with a balloon mounted on the distal tip.It has a radiopaque tip and two radiopaque markers beneath the balloon that define the working length.The lumen labeled with the rated burst pressure (xx atm) is for balloon inflation.The other lumen allows the catheter to track over a 0.038¿ (.97mm) diameter guidewire and can be used for monitoring of pressure or the infusion of medication and/or contrast medium.Each balloon inflates to a stated diameter and length at a specific pressure, typically 10 atm.The x-force® balloon dilation catheter comes packaged with a refolding tool and is available with or without an inflation device.It comes sterile and is for single use only.Indications for use the x-force® balloon dilation catheter is recommended for use in the dilation of the urinary tract.Contraindications do not use the x-force® balloon dilation catheter in the presence of conditions, which create unacceptable risk during the dilation of the urinary tract.Warnings ¿ do not exceed the recommended rated burst pressure (rbp) for this device.Balloon rupture may occur if the rbp rating is exceeded.Please refer to the device label for the recommended maximum rated burst pressure.Extreme caution should be used when considering dilation of irregular, not-compliant tissue or in the presence of certain calculi.¿ this is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Precautions caution: federal law (u.S.A.) restricts this device to sale by or on the order of a physician.Only a physician who has an understanding of the clinical applications, technical principles and associated risks of balloon dilation within the urinary tract should use this device.After use, the x-force® balloon dilation catheter and/or accessories may be considered a potential biohazard.Handle and dispose of in accordance with medical practice and applicable laws and regulations.Inspection prior to use the x-force® balloon dilation catheter is a sterile, single use device.Carefully inspect the catheter and the sterile packaging for signs of damage that may have occurred during shipment.Do not use the product if damage is evident.Preparation of the catheter all x-force® balloon dilation catheters contain air in the balloon lumen.The air must be removed to allow liquid to fill the balloon when it is inflated.1.Remove the protective sheath from the balloon.2.Attach the inflation device to the connector on the balloon lumen.3.Open the stopcock and draw back on the inflation device to remove the air from the balloon catheter.4.Close the stopcock, remove the inflation device, depress the plunger to remove any air and reattach to the balloon catheter.5.Repeat steps 3-4 until all air is removed from the balloon lumen.Catheter insertion 1.Introduce the catheter carefully over a 0.038¿ (.97mm) guidewire and place it in the area that needs to be dilated.Use the radiopaque markers to aid in proper positioning.Note: dilation procedures should be conducted under fluoroscopic guidance with appropriate x-ray equipment or direct vision.Caution: do not advance or withdraw the catheter or guidewire against any significant resistance.The cause of the resistance must be determined fluoroscopically and remedial action taken.Inflating the balloon catheter 1.Fill the inflation device (eagle¿ inflation device) with sterile diluted contrast medium.2.Attach the inflation device to the balloon lumen.3.Inflate the balloon note: do not use air or any gaseous substances as a balloon inflation medium always use sterile liquid media.Note: the use of an inflation device with a pressure gauge is highly recommended to make sure adequate pressure is applied and the rated burst pressure of the balloon is not exceeded.Caution: if loss of pressure in the balloon occurs or the balloon ruptures, immediately stop the procedure, deflate the balloon and remove carefully.Do not attempt to re-inflate the balloon.If after inspection it is noted pieces of the balloon are missing, check for and endoscopically retrieve the fragments when possible.Warning: do not exceed the recommended rated burst pressure (rbp) for this device.Balloon rupture may occur if the rbp rating is exceeded.Please refer to the device label for the recommended maximum rated burst pressure.Extreme caution should be used when considering dilation of irregular, not-compliant tissue or in the presence of certain calculi.Deflating the balloon catheter deflate the balloon using an inflation device.Since deflation times vary based on balloon sizes and shaft lengths, check fluoroscopically to confirm deflation before attempting to withdraw.1.Attach the inflation device (eagle¿ inflation device) to the balloon catheter and remove the solution from the balloon by pulling back on the inflation device.2.Remove the inflation device from the balloon catheter.This will allow ambient pressure to enter, relaxing the balloon.3.Gently withdraw the catheter.Use of a gentle counterclockwise twisting motion is recommended when removing the catheter.Caution: if resistance is felt when removing either the catheter or the guidewire from the introducer sheath, stop and consider removing them as a single unit to prevent damage to the product.Applying excessive force to the catheter can result in tip breakage or balloon separation.Using the refolding tool the balloon catheter is supplied with a refolding tool.This aids in preparing the balloon for another insertion during the procedure.To use: 1.Manually compress the balloon.2.Position the refolding tool at one end of the balloon.3.Twist the refolding tool counter clockwise and push down on the balloon until the refolding tool traverses the entire length of the balloon.4.Once the balloon is folded, remove the refolding tool and proceed with the procedure.Dispose of properly.Storage store the balloon dilation catheter in a cool, dry, dark place.Do not store near radiation or ultraviolet light sources as these may damage product materials.¿ correction: b h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that the customer opened the package of the ureteral balloon dilation catheter.And found that it could not be pressurized after normal use, and then took out the balloon for testing and found that the tip was leaking.The user experienced prolonged operation time and has been exposed to harmful substances, blood, or bodily fluids.
|