| Model Number |
FIBERTAK RC, DOUBLOAD TAPE BL/W, BLK/W |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force during tensioning and/or improper bone prep.
|
| |
|
Event or Problem Description
|
|
It was reported that during a pectoral tear shoulder surgery the device tore out.No part of the device broke off.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a different device by switching to a cork screw procedure.It was not necessary to do a second surgery.
|
| |
|