| Model Number |
VTICM512.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no similar complaint was reported for units within the same lot.Type of investigation 3331 - device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted and removed a 12.6mm vticm5 12.6 implantable collamer lens -8.00/2.50/86 (sphere/cylinder/axis) into the patient's left eye (os) on (b)(6) 2023.The lens was reported as having low vaulting without rotation.The lens was intraoperatively replaced with a longer length lens.This resolved the problem.
|
| |
|