Model Number OF1024038S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Local Reaction (2035)
Date of Event 04/01/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Company was notified of a scheduled removal surgery following localized pain in the operated right foot 8 months after implantation.The removal surgery was completed successfully.Patient also had four implants removed from the left foot a month earlier due to similar symptoms.An investigation was performed including internal analysis of records, surgeon feedback, and analysis of symptoms.There is no indication of a design, manufacturing, sterilization, or process issue affecting implant safety or effectiveness.There were no issues during the primary procedure.Patient post op recovery was uneventful, and bone healing and fusion of the operated sites were achieved.Based on the similarity of the symptoms exhibited in the right foot and in the left foot it was concluded that the root cause is likely the same.The investigation of the left foot event ruled out infection based on imaging, lab work, negative cultures and timing of symptoms, and concluded that the root cause of the pain could not be determined but is likely the outcome of the patient experiencing a local reaction to one of the implant's materials or related to an unreported underlying medical condition.Instructions for use list an allergic reaction as a potential adverse effect to implantation of foreign material.Company continues to monitor these events as part of post market activities.Additional reports relating to this event: # 3014554088-2024-00007/9/10 & 3014323288-2024-00007/8/9/10.Additional reports relating to the left foot of the same patient: # 3014554088-2024-00003/4/5/6 & 3014323288-2024-00003/4/5/6.
 
Event or Problem Description
Pain in the right foot following a bunion deformity correction.Four implants were removed 8 months post-op.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSIOFIBER COMPRESSION SCREW 4.0X38MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
OSSIO LTD.
8 hatochen st.
caesarea, 30798 61
IS  3079861
Manufacturer (Section G)
OSSIO LTD.
8 hatochen st.
caesarea, 30798 61
IS   3079861
Manufacturer Contact
taly lindner
8 hatochen st.
caesarea, 30798-61
IS   3079861
MDR Report Key19255192
Report Number3014554088-2024-00008
Device Sequence Number15291668
Product Code HWC
UDI-Device Identifier07290017630342
UDI-Public(01)07290017630342(17)240102(11)220103(10)OF01517
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K193660
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/02/2024
Device Model NumberOF1024038S
Device Catalogue NumberOF1024038S
Device Lot NumberOF01517
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2024
Initial Report FDA Received Date05/06/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OSSIOFIBER COMPRESSION SCREW 3.5X20MM; OSSIOFIBER COMPRESSION SCREW 4.0X32MM; OSSIOFIBER COMPRESSION STAPLE SYSTEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age19 YR
Patient SexFemale
-
-