• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD. POCKET PAIN PRO TENS UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD. POCKET PAIN PRO TENS UNIT Back to Search Results
Model Number PM400
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code Partial thickness (Second Degree) Burn (2694)
Date of Event 04/23/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The consumer sent an email stating that he purchased a tens unit from a cvs in phoenix, az.He used the unit once as instructed and was told by the va hospital that he received second degree burns.During a follow-up call, the consumer stated his wife bought the unit for his bursitis.He stated it was amazing what it did when he first used the unit for his bursitis.He stated that after a while he removed the unit and he saw a burn so he went to the va hospital, and they gave him medicine for a second degree burn.He stated he was very excited about the unit but hesitant to use it again.He stated he was walking or sitting down when the burn happened.He was wearing stretchy pants over them like sweatpants.The pads were placed 6-7 inches apart from one another.There were no other devices nearby.He stated he used the unit for one round of 15 minutes.He was the only user.He stated the event took place on april 23rd around 5-6 pm.He went to the va hospital on (b)(6).He was using the original pads.He stated the burn was on his right hip and buttocks right where the hip bones meet.He stated he was not allergic to anything, but he remembers when he was young that inner tubes would irritate his skin.He stated he had sensitive skin.He stated he used the unit after showering and using body wash.He made sure the area was dry.He peeled off the label and used both pads, and there was no oil or lotion on the skin.He was prescribed ointment to put on the burn and had no other treatments done during the visit.He stated he was getting a shot of cortisone for his bursitis instead.He agreed to send the unit back for inspection and requested a refund.The consumer called back on may 1st, 2024 to speak to the call center representative.She was unavailable at the time so he hung up.She called back, but he did not answer.
 
Additional Manufacturer Narrative
A root cause has not been determined.It has not been confirmed if the device caused or contributed to the reported incident; however, due to the customer reporting having burns this medwatch is being filed.The consumer was offered a refund.The product instruction manual includes following warnings: - the mere existence of pain functions as a very important warning telling us that something is wrong.Therefore, if you suffer from any serious illness, consult your physician in order to confirm that it is advisable for you to use this unit.- if you experience any skin irritation or redness after a session, do not continue stimulation in that area of the skin.- place pads on either side of the pain, not directly on the pain.Possible adverse reactions: - you should stop using the unit and consult with your physician if you experience adverse reactions from the unit.- you may experience skin irritation and burns beneath the stimulation electrodes applied to your skin.
 
Event or Problem Description
The consumer sent an email stating that he purchased a tens unit from a cvs in phoenix, az.He used the unit once as instructed and was told by the va hospital that he received second degree burns.During a follow-up call, the consumer stated his wife bought the unit for his bursitis.He stated it was amazing what it did when he first used the unit for his bursitis.He stated that after a while he removed the unit and he saw a burn so he went to the va hospital, and they gave him medicine for a second degree burn.He stated he was very excited about the unit but hesitant to use it again.He stated he was walking or sitting down when the burn happened.He was wearing stretchy pants over them like sweatpants.The pads were placed 6-7 inches apart from one another.There were no other devices nearby.He stated he used the unit for one round of 15 minutes.He was the only user.He stated the event took place on (b)(6) 2024 around 5-6 pm.He went to the va hospital on (b)(6) 2024.He was using the original pads.He stated the burn was on his right hip and buttocks right where the hip bones meet.He stated he was not allergic to anything, but he remembers when he was young that inner tubes would irritate his skin.He stated he had sensitive skin.He stated he used the unit after showering and using body wash.He made sure the area was dry.He peeled off the label and used both pads, and there was no oil or lotion on the skin.He was prescribed ointment to put on the burn and had no other treatments done during the visit.He stated he was getting a shot of cortisone for his bursitis instead.He agreed to send the unit back for inspection and requested a refund.The consumer called back on may 1st, 2024 to speak to the call center representative.She was unavailable at the time so he hung up.She called back, but he did not answer.
 
Additional Manufacturer Narrative
A root cause has not been determined.It has not been confirmed if the device caused or contributed to the reported incident; however, due to the customer reporting having burns this medwatch is being filed.The consumer was offered a refund.The product instruction manual includes following warnings: - the mere existence of pain functions as a very important warning telling us that something is wrong.Therefore, if you suffer from any serious illness, consult your physician in order to confirm that it is advisable for you to use this unit.- if you experience any skin irritation or redness after a session, do not continue stimulation in that area of the skin.- place pads on either side of the pain, not directly on the pain.Possible adverse reactions: - you should stop using the unit and consult with your physician if you experience adverse reactions from the unit.- you may experience skin irritation and burns beneath the stimulation electrodes applied to your skin.************************************************************************************************************************************* updated information in section e1, f3-5, f7.Importer and initial report contact information was switched.
 
Event or Problem Description
The consumer sent an email stating that he purchased a tens unit from a (b)(6).He used the unit once as instructed and was told by the (b)(6) hospital that he received second degree burns.During a follow-up call, the consumer stated his wife bought the unit for his bursitis.He stated it was amazing what it did when he first used the unit for his bursitis.He stated that after a while he removed the unit and he saw a burn so he went to the (b)(6) hospital, and they gave him medicine for a second degree burn.He stated he was very excited about the unit but hesitant to use it again.He stated he was walking or sitting down when the burn happened.He was wearing stretchy pants over them like sweatpants.The pads were placed 6-7 inches apart from one another.There were no other devices nearby.He stated he used the unit for one round of 15 minutes.He was the only user.He stated the event took place on (b)(6) 2024 around 5-6 pm.He went to the (b)(6) hospital on (b)(6).He was using the original pads.He stated the burn was on his right hip and buttocks right where the hip bones meet.He stated he was not allergic to anything, but he remembers when he was young that inner tubes would irritate his skin.He stated he had sensitive skin.He stated he used the unit after showering and using body wash.He made sure the area was dry.He peeled off the label and used both pads, and there was no oil or lotion on the skin.He was prescribed ointment to put on the burn and had no other treatments done during the visit.He stated he was getting a shot of cortisone for his bursitis instead.He agreed to send the unit back for inspection and requested a refund.The consumer called back on (b)(6) 2024 to speak to the call center representative.She was unavailable at the time so he hung up.She called back, but he did not answer.
 
Additional Manufacturer Narrative
A root cause has not been determined.It has not been confirmed if the device caused or contributed to the reported incident; however, due to the customer reporting having burns this medwatch is being filed.The consumer was offered a refund.The product instruction manual includes following warnings: the mere existence of pain functions as a very important warning telling us that something is wrong.Therefore, if you suffer from any serious illness, consult your physician in order to confirm that it is advisable for you to use this unit.If you experience any skin irritation or redness after a session, do not continue stimulation in that area of the skin.Place pads on either side of the pain, not directly on the pain.Possible adverse reactions: you should stop using the unit and consult with your physician if you experience adverse reactions from the unit.You may experience skin irritation and burns beneath the stimulation electrodes applied to your skin.Updated information in section e1, f3-5, f7.Importer and initial report contact information was switched.Updated information in section f7 and included additional follow-up information from the manufacturer below in section h11.The device was not returned to the distributor/manufacturer for investigation.The investigation was closed on 7/25/2024.Here is the summary of the manufacturer device investigation: the manufacturer reviewed the qa test data, risk analysis document and manufacturer shipping inspection.No issue/problem was noted based on the data reviewed.There was one similar event reported for the same model in the past.However, no abnormal trend was identified for similar models post market.Unit was not returned by consumer.The device was not received for evaluation; therefore, a device analysis could not be completed.No further investigation required.There has not been any additional communication from this consumer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POCKET PAIN PRO TENS UNIT
Common Device Name
TENS
Manufacturer (Section D)
OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD.
no. 28 vsip ii, st. 2,
vietnam-singapore indus. park
hoa phu thu dau mot city, 82000 0
VM  820000
MDR Report Key19255487
Report Number1450057-2024-00002
Device Sequence Number6381272
Product Code NUH
UDI-Device Identifier00073796634001
UDI-Public(01)00073796634001
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/06/2024
Report Date (Section F) 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberPM400
Device Catalogue NumberPM400
Device Lot Number202308V
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date04/30/2024
Device Age8 MO
Event Location Home
Date Report to Manufacturer05/06/2024
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date05/06/2024
Supplement Report FDA Received Date05/07/2024
08/06/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age55 YR
Patient SexUnknown
Date Report Sent to FDA05/06/2024
-
-