BIOTRONIK SE & CO. KG ETRINSA 8 HF-T; CRT-P Back to Search Results
Model Number 394919
Medical Device Problem Code Ambient Noise Problem (2877)
Health Effect - Clinical Codes Bradycardia (1751); Loss of consciousness (2418)
Date of Event 04/26/2024
Type of Reportable Event Serious Injury
Event or Problem Description
While nurse was changing dressing, patient passed out and had no pulse, then woke up on their own with pulse around 40bpm (pacemaker programmed to 60bpm).Emt also reported 40bpm pulse.Hmsc shows no unusual behavior or episodes on that day.Pacing percentage on that day reported 100 percent.Discussed possibilities of noise from lead or external emi interfering with device.Recommended device evaluation with isometric testing.Device remains implanted.
 
Additional Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETRINSA 8 HF-T
Common Device Name
CRT-P
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key19255993
Report Number1028232-2024-02518
Device Sequence Number2608216
Product Code NKE
UDI-Device Identifier04035479131777
UDI-Public(01)04035479131777(17)180831
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P950037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2018
Device Model Number394919
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2024
Supplement Date Received by Manufacturer05/08/2024
Initial Report FDA Received Date05/06/2024
Supplement Report FDA Received Date05/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/10/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexMale
-
-