Catalog Number VBJR070502A
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.H6 - code c19: review of device manufacturing record history confirmed device met pre-release specifications.H6 - code c21: returned specimen is currently undergoing investigation.Results are pending.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
The following was reported to gore: during an emergency procedure for a patient who had a stab wound in the left axillary artery, a gore® viabahn® endoprosthesis with heparin bioactive surface (viabahn device) did not deploy.As reported, access was made from the common femoral artery.A 7fr cook sheath was used to advance the viabahn device over a 0.035 amplatz wire.When the viabahn device was placed, the deployment line was pulled but the line was stuck.After several attempts, the constrained viabahn device was removed.A new viabahn device was advanced and deployed at the intended treatment site with no issues.The patient is reported to be recovering well following the procedure.The inability to deploy device did not adversely impact the patient as physician was able to use another device.
 
Additional Manufacturer Narrative
H6: code c20: returned specimen underwent investigation.Results were: the vsx device was returned for evaluation with the outer zipper very slightly deployed (1.5 mm) and the endoprosthesis still constrained on the delivery catheter.Deployment of the returned vsx device endoprosthesis was not able to continue when pulling the deployment line from the hub or at the endoprosthesis.The cause of the stuck deployment line could not be established during evaluation.Procedural and benchtop deployment of the device can be impacted by different factors including but not limited to zipper integrity, delivery system support or stiffness, or presence of dried fluid on the device.The investigation could not confirm the cause of the reported hazardous situation nor assign a primary device failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
genevieve begay
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key19256003
Report Number2017233-2024-04897
Device Sequence Number2685358
Product Code NIP
UDI-Device Identifier00733132623761
UDI-Public00733132623761
Combination Product (Y/N)Y
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P040037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/21/2026
Device Catalogue NumberVBJR070502A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2024
Supplement Date Received by Manufacturer06/10/2024
Initial Report FDA Received Date05/06/2024
Supplement Report FDA Received Date06/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age24 YR
Patient SexMale
-
-