MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT TIB IMP CRC INSERT SZ 4.5, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT TIB IMP CRC INSERT SZ 4.5, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-022-51-4511
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Numbness (2415); Joint Laxity (4526); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via legal documentation that approximately 35 months after a left total knee replacement procedure, the patient may require future revision surgery for polyethylene wear.The patient is currently claiming the following complication/injuries as a result of their implant: pain, instability, swelling, and numbness.No additional information is available.
 
Additional Manufacturer Narrative
D10: concomitant devices: serial number , item number and full description: (b)(6), a10012 - gps implant kit v2; (b)(6), 02-020-14-0245 - truliant cr por fem cr por left sz 4.5; (b)(6), 201-78-98 - 2 pk, schanz pin, 4mm x 130mm; (b)(6), 02-022-55-4545 - truliant por tib tray size 4.5f/4.5t, (b)(6), 201-78-15 - holding pin mini sharp point 4 pk; (b)(6), 521-78-32 - threaded pin size 3.0 collarless 2pk; (b)(6), 521-78-32 - threaded pin size 3.0 collarless 2pk; (b)(6), 521-78-23 - threaded pin size 2.3 collared 2pk.The device was not returned for evaluation and no photographs or x-rays were provided for review; therefore the reported event cannot be confirmed through analysis.No operative notes were provided; therefore, a review of device usage and technique could not be performed.No information concerning patient conditions, co-morbidities, or other health or anatomical concerns was provided and it is therefore unknown if patient-related factors may have caused or contributed to the reported event.The cause of the reported event cannot be determined based on the information made available.Should additional, relevant information become available, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: g3, h6.The following sections were corrected: h6.Mdr section codes updated/corrected: a, b, e, f.The reason for the reported event cannot be confirmed from the information provided but may be due to prosthesis wear and instability or inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT TIB IMP CRC INSERT SZ 4.5, 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key19258384
Report Number1038671-2024-01077
Device Sequence Number20486735
Product Code JWH
UDI-Device Identifier10885862307828
UDI-Public10885862307828
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K171045
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number02-022-51-4511
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2024
Supplement Date Received by Manufacturer12/10/2024
Initial Report FDA Received Date05/07/2024
Supplement Report FDA Received Date12/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-