This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, g4, g7, h1, h2, h3, h6, h10.Initial medwatch was submitted with a notification date of apr 15, 2024 in g3 and submitted on may 7, 2024; however the correct notification date is mar 20, 2024.H6: proposed component (annex g) code is mechanical (g04) - provisional bottom.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows signs of repeated use and confirming complaint post fractured off.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.The summary of investigations for complaints associated with the tibial articular surface provisional (tasp) fracture.Fractured tasps were analyzed by optical microscopy revealing hackle marks and river lines in the case of bending overload; and hackle marks, river lines and striations were found in the case of low cycle fatigue culminating in a bending overload failure type.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|