Qn#(b)(4).The customer returned two units 5-10116 et tube,cf,8.0 for investigation.Two representative samples were returned.An actual sample was not returned.The representative samples were visually examined with and without magnification.Visual examination of the returned devices revealed that the samples appear typical.No defects or anomalies were observed.A functional inspection was performed to attempt to inflate and deflate the balloon cuff on the returned et tubes.A lab inventory syringe filled with air was connected to the valve of the pilot balloon of one of the returned samples.Upon injection of air, both the pilot and cuff immediately inflated.The syringe was then removed and the cuff and pilot balloon were inspected for leaks.No leaks were detected.The syringe was then reattached in order to deflate the balloons.Both the pilot balloon and balloon cuff were able to deflate.The sample was submerged underwater to test for any leaks.However, no leaks were detected.This process was repeated for the other sample with the same results.No issues were found with the returned samples as they were both able to properly inflate and deflate.A device history record review was performed and no relevant findings were identified.The states, "the tracheal tube cuff inflation system, including the valve, should be checked prior to intubation.If a malfunction is a detected in any part of the system, the tube should not be used." "care should be taken to avoid damaging the thin-walled cuffs during intubation.If cuff is damaged, the tube should not be used." the ifu also states, "cuff pressure should be monitored routinely to assure that an adequate tracheal seal is maintained and that the cuff has not become over-inflated.The tracheal tube cuff should be slowly inflated with the minimum amount of air required to provide an effective tracheal seal.Do not inflate with a measured volume of air or by "feel" of pressure in the syringe since little resistance should be felt during inflation." the reported complaint could not be confirmed since the actual sample was not returned.The customer returned two representative samples in place of the actual sample that resulted in the complaint issue.Upon functional inspection, no problems were found with the returned devices as they were both able to properly inflate and deflate.No leaks were found with the samples.A device history record review was performed on the device with no evidence to suggest a manufacturing related cause.The probable cause of this complaint issue could not be determined without the actual device.
|