Catalog Number 5-10116
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported "ett tube did not inflate during procedure".No patient harm or injury.No medical intervention required.A new endotracheal tube was placed.The patient's condition was reportedly "stable before and after equipment malfunction".
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event or Problem Description
It was reported "ett tube did not inflate during procedure".No patient harm or injury.No medical intervention required.A new endotracheal tube was placed.The patient's condition was reportedly "stable before and after equipment malfunction".
 
Additional Manufacturer Narrative
Qn#(b)(4).The customer returned two units 5-10116 et tube,cf,8.0 for investigation.Two representative samples were returned.An actual sample was not returned.The representative samples were visually examined with and without magnification.Visual examination of the returned devices revealed that the samples appear typical.No defects or anomalies were observed.A functional inspection was performed to attempt to inflate and deflate the balloon cuff on the returned et tubes.A lab inventory syringe filled with air was connected to the valve of the pilot balloon of one of the returned samples.Upon injection of air, both the pilot and cuff immediately inflated.The syringe was then removed and the cuff and pilot balloon were inspected for leaks.No leaks were detected.The syringe was then reattached in order to deflate the balloons.Both the pilot balloon and balloon cuff were able to deflate.The sample was submerged underwater to test for any leaks.However, no leaks were detected.This process was repeated for the other sample with the same results.No issues were found with the returned samples as they were both able to properly inflate and deflate.A device history record review was performed and no relevant findings were identified.The states, "the tracheal tube cuff inflation system, including the valve, should be checked prior to intubation.If a malfunction is a detected in any part of the system, the tube should not be used." "care should be taken to avoid damaging the thin-walled cuffs during intubation.If cuff is damaged, the tube should not be used." the ifu also states, "cuff pressure should be monitored routinely to assure that an adequate tracheal seal is maintained and that the cuff has not become over-inflated.The tracheal tube cuff should be slowly inflated with the minimum amount of air required to provide an effective tracheal seal.Do not inflate with a measured volume of air or by "feel" of pressure in the syringe since little resistance should be felt during inflation." the reported complaint could not be confirmed since the actual sample was not returned.The customer returned two representative samples in place of the actual sample that resulted in the complaint issue.Upon functional inspection, no problems were found with the returned devices as they were both able to properly inflate and deflate.No leaks were found with the samples.A device history record review was performed on the device with no evidence to suggest a manufacturing related cause.The probable cause of this complaint issue could not be determined without the actual device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ET TUBE,CF,8.0
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key19260390
Report Number3003898360-2024-00431
Device Sequence Number17812491
Product Code BTR
UDI-Device Identifier04026704396825
UDI-Public04026704396825
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K822082
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number5-10116
Device Lot Number73G2300459
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2024
Supplement Date Received by Manufacturer05/10/2024
Initial Report FDA Received Date05/07/2024
Supplement Report FDA Received Date05/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED
Patient SexUnknown
-
-