MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3543507 |
| Medical Device Problem Codes |
Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/15/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made. as such, the investigation will be closed. if the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803. this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date. this report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report. if certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: capsular contracture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation revision surgery with 350cc mentor memorygel breast implants.Post-operatively, the patient experienced a rupture of her left prosthesis and baker grade iv capsular contracture of the right breast, which were confirmed via mammogram.As a result, the patient underwent bilateral removal surgery on (b)(6) 2024.This report is for the right implant.Refer to manufacturing report number 1645337-2024-05355 for the contralateral event.
|
| |
|
Additional Manufacturer Narrative
|
|
On may 08, 2024, mentor received additional information.Results of magnetic resonance imaging indicated that the patient experienced bilateral rupture of her prostheses.It was reported that the patient experienced swelling in her left breast.A minimal amount of subcapsular fluid was also observed in the right breast.On may 23, 2024, mentor received the device for evaluation.The analysis has begun but is not complete at this time. when the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On may 30, 2024, mentor completed a visual inspection of the returned device.Device evaluation summary: visual analysis of the returned sample revealed that the breast implant had an area of siltex cracking on the posterior view.In addition, a tear was noted within the siltex cracking measuring approximately 3.2 cm.The evaluation determined that the possible cause of the rupture was consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsules, and effusion to pathology for examination.Manufacturer¿s reference number: (b)(4).
|
| |
|