|
This report is being supplemented to provide additional information based on the approved final investigation.It has been one (1) year since the subject device was manufactured.A review of the device history record found no deviations or non-conformities that could have caused or contributed to the reported issue.Based on the results of the investigation, the cutting knife is inferred to have been fractured; however, the root cause could not be identified.During the tissue incision, discharge occurred due to the output setting being high, the coagulation mode being used, and the energization time being prolonged.The high temperature produced by the discharge led to a deterioration in the strength of the cutting knife.The cutting knife, which had reduced strength, was subjected to a tensile load, causing it to thin and fracture at the base.Consequently, the fractured area of the cutting knife assumed a tapered shape.The gloss observed on the fracture surface is concluded to have resulted from friction with another object.The event can be prevented by following the instructions for use which state: (drawing no.Gk9165 revision no.9) during treatment, always ensure that the slider slides on the handle smoothly and that the electrosurgical knife observed in the endoscopic image is normal.Should deformation or break of the cutting knife be detected during use, immediately shut off the power supply, discontinue the procedure, pull the slider and withdraw the endoscope from the patient with the cutting knife retreated in the coated outer tube.Do not continue using an abnormal electrosurgical knife to prevent perforation or bleeding.If the cutting knife is detached, be sure to collect it using a grasping forceps.Always operate the electrosurgical unit at the minimum output level and for the minimum time necessary to successfully complete the procedures.Excessive output level and time may result in patient injury, such as perforation, bleeding, or mucous membrane damage.Application of high-voltage waveforms for extended periods increases the likelihood that the cutting knife may break.When a high-voltage waveform has to be used, minimize the duration of current application.Electrosurgical unit: voltage intensities of various waveforms.1: soft coagulation < cut / pulse cut < forced coagulation.Do not activate output when the metal portion at the distal end of the endoscope is too close to or in contact with the cutting knife.This could burn the tissue and/or damage the endoscope.If the cutting knife is used in a high-humidity situation, it may be damaged by melting or elongation.Do not use this instrument and a cord with an output higher than the rated high-frequency voltage given in the tables in section 8.2, ¿specifications¿.This could cause patient, operator, or assistant injury, such as thermal injury.It could also damage the endoscope, instrument, and/or a cord.8.2.Rated high-frequency voltage cut: 1600 vp (3200 vp-p) coag: 2900 vp (5800vp-p) be careful not to use excessive force when removing tissue attached to the cutting knife.When the distal end is subjected to excessive force, for example, when forcibly scraping the cutting knife with tweezers, etc.Or when extending and retracting the cutting knife abruptly and continuously, it may break the cutting knife.Olympus will continue to monitor field performance for this device.
|