| Model Number |
CLV-190 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the light source had connector issues.It works with some scopes, but not all.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the manufacturer¿s investigation.A device evaluation, a review of the device history record (dhr), as well as a historical trending analysis were conducted during this investigation.The subject device was returned, evaluated, and the user¿s report was not confirmed.The review of the dhr did not find any abnormalities or anomalies identified during production.The device met all specifications upon release.Based on the results of the investigation and the condition of the returned device, a definitive root cause for the reported failure mode could not be determined.However, investigators believe it¿s likely that the reported failure was the result of a faulty latch mechanism on the output socket.Olympus will continue to monitor the field performance of this device.
|
| |
|