EXACTECH, INC. OPTETRAK CLASSIC AND TRULIANT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
200-24-09 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
01/25/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: concomitant devices: (b)(6); 200-04-45 - cemented finned tib.Tra sz 4f/5t; (b)(6); 200-02-35 - three peg patella 35mm; (b)(6); 230-03-04 - optetrak asy,cr cemented femoral, sz 4, the device was not returned for evaluation and no photographs or x-rays were provided for review; therefore the reported event cannot be confirmed through analysis.No operative notes were provided; therefore, a review of device usage and technique could not be performed.No information concerning patient conditions, co-morbidities, or other health or anatomical concerns was provided and it is therefore unknown if patient-related factors may have caused or contributed to the reported event.The cause of the reported event cannot be determined based on the information made available.Should additional, relevant information become available, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that approximately 108 months after a bilateral knee replacement procedure, the patient underwent a revision procedure to address polyethylene wear and other unspecified causes.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
D6b: explant date removed.H6: corrected health effect clinical code, health effect impact code, type of investigation.
|
| |
|