| Catalog Number |
71751135 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an internal fixation surgery, the humeral small hammer fragmented when the nail was being impacted.The procedure was resumed, without any delay, using a competitors device.Even though it is unknown if broken pieces fell into the patient, no injury was reported as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
H11- corrected data.B5- describe event or problem.H6- evaluation codes.
|
| |
|
Event or Problem Description
|
|
It was reported that during an internal fixation surgery, the humeral small hammer fragmented when the nail was being impacted.No pieces fell inside the patient and no injury was reported as consequence of this problem.The procedure was resumed, without any delay, using a competitor's device.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, per complaint details, it was noticed during a total knee arthroplasty that the hammer fragmented during an internal fixation surgery.Reportedly, no pieces fell inside the patient and the procedure was resumed using a competitor device without delay.As of the date of this medical investigation, the requested clinical documentation has not been provided and no clinical root cause can be concluded to have contributed to the reported event.Reportedly the patient was not injured as a consequence of this problem.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|