Model Number SUTURE ANCHOR, MICRO CORKSCREW FT
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint confirmed.Two unpackaged ar-1318ft-40 (no batch indicators present) were received for investigation.Visual inspection identified that the distal tips of both inserters were severely warped.The observed condition is consistent with prying/leveraging the inserters within the anchors during use.
 
Event or Problem Description
On 10/8/2021, it was reported by an arthrex employee via email that (2) ar-1317ft nano corkscrew and (2) ar-1318ft-40 micro suture anchor corkscrew tips bent while insertion.This was discovered during a procedure on (b)(6) 2021, when surgeon made a pilot hole, ar-1317ft tip bent and the bone quality was hard.This was attempted again with another nano corkscrew and the driver tip bent when it was about 80% inserted.The same happened when ar-1318ft-40 screw was inserted.Finally, surgeon was able to insert and suture one nano and one micro corkscrew to complete case.Additional information received 10/11/2021 surgeon tried to use another ar-1318ft-40 a second time and tip was also bent.There were no broken pieces inside patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTURE ANCHOR, MICRO CORKSCREW FT
Common Device Name
BONE FIXATION SCREW/PLATE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19265955
Report Number1220246-2024-02925
Device Sequence Number12620131
Product Code HWC
UDI-Device Identifier00888867008526
UDI-Public00888867008526
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112237
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2025
Device Model NumberSUTURE ANCHOR, MICRO CORKSCREW FT
Device Catalogue NumberAR-1318FT-40
Device Lot Number12400290
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2021
Initial Report FDA Received Date05/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-