• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. ACL TIGHTROPE RT W/ DEPLOYING SUTURE; SMOOTH FIXATION PIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. ACL TIGHTROPE RT W/ DEPLOYING SUTURE; SMOOTH FIXATION PIN Back to Search Results
Model Number ACL TIGHTROPE RT W/ DEPLOYING SUTURE
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an anterior cruciate ligament surgery the device could not be tightened.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint not confirmed.-upon visual inspection no problem found with the ar-1588rt-j.Functional testing has shown that the device can be tightened.No problems found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACL TIGHTROPE RT W/ DEPLOYING SUTURE
Common Device Name
SMOOTH FIXATION PIN
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19267679
Report Number1220246-2024-02945
Device Sequence Number7430431
Product Code HTY
UDI-Device Identifier00888867095533
UDI-Public00888867095533
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K112990
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberACL TIGHTROPE RT W/ DEPLOYING SUTURE
Device Catalogue NumberAR-1588RT-J
Device Lot Number14983756
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2022
Initial Report FDA Received Date05/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-