|
It was reported that male external catheter was leaking and causing issue with their wounds.They went to wound clinic on 21mar to get their wounds cleaned as the leakage was causing infections.Per sample received on 22apr2024, the customer had returned additional sample with batch number juhy1564.Per sample form received on 22apr2024, it was reported that the patient tried three of the catheters from lot#juhy1564 and one from lot#juhy0636.All four leaked overnight and patient was drenched in urine causing to the wounds and went into the hospital from 29mar2024 to 09apr2024, found out wound infected from the urine.Patient was treated on intravenous toxin antibiotic for 10 days to clear bacterial infection.
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that male external catheter was leaking and causing issue with their wounds.They went to wound clinic on 21mar to get their wounds cleaned as the leakage was causing infections.Per sample received on 22apr2024, the customer had returned additional sample with batch number juhy1564.Per sample form received on 22apr2024, it was reported that the patient tried three of the catheters from lot#juhy1564 and one from lot#juhy0636.All four leaked overnight and patient was drenched in urine causing to the wounds and went into the hospital from (b)(6) 2024, found out wound infected from the urine.Patient was treated on intravenous toxin antibiotic for 10 days to clear bacterial infection.
|
|
The reported event is inconclusive as representative lot samples evaluated are within specification but condition of the used original sample is unknown.Visual: received one (1) ultraflex male external catheter in closed packaging.No blockage, holes, rips or tears present.Therefore, product meets specifications.The exact root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be viscometer failure.However, there was insufficient information to confirm this potential root cause.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: ¿ultraflex¿ self-adhering male external catheter.Spirit¿ style 1 hydrocolloid adhesive sheath.Intended use: the self-adhering male external catheter is used for the drainage of urine.The catheter is applied by the patient or caregiver.Description / indication: the self-adhering male external catheter is designed for the management of adult male urinary incontinence.Contraindication: do not use on irritated or compromised skin.Warning: reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/ or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury or illness of the patient.Precaution: do not use if allergic reaction occurs or if patient has known allergies to device components.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Follow directions to remove if experiencing swelling, numbness, discomfort, pain, discoloration, or abnormal appearance.Note: if experiencing problems with use of the device, please consult your healthcare professional for assistance.Directions to apply: 1) verify correct size prior to use.2) trim pubic hair if necessary.3) wash penis with mild soap and warm water.Dry thoroughly.Wear time may be reduced if skin is not dry or cream/oil is used.4) open package at perforation.Remove catheter from plastic insert, if present.5) place rolled end over the end of the penis, leaving a small space between the end of the penis and the cone of the catheter.6) unroll the catheter over penis.7) gently squeeze the catheter to properly seal adhesive to the skin.If possible, avoid leaving a rolled ¿collar¿ around the base of the penis.Important: wear time may be reduced if adhesive is not properly sealed to the skin.8) connect the catheter to a drainage system.Make sure to check that connections are secure before use.Directions to remove: 1) ensure drainage bag is empty.2) disconnect catheter from the drainage system.3) gently roll the catheter forward and off the penis.If necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive.Disposal: after removal, dispose by placing the used catheter into a waste container.¿ h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|