• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. A.T.S. 4000TS TOURNIQUET W/HOSES AND LOP SENSOR (PERSONALIZED PRESSURE®); TOURNIQUET, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. A.T.S. 4000TS TOURNIQUET W/HOSES AND LOP SENSOR (PERSONALIZED PRESSURE®); TOURNIQUET, PNEUMATIC Back to Search Results
Model Number N/A
Medical Device Problem Code Measurement System Incompatibility (2982)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/03/2024
Type of Reportable Event Malfunction
Event or Problem Description
During the calibration process, they applied pressure to the ats, however the pressure source was fully compressed with little to no back pressure from the ats.The device passes calibration on the blue side (cuff side), however going down the red side (reservoir side), the multiple syringes that we used as a pressure source are not holding pressure on the 250 test or more.The pressure setting did not match the actual pressure it was set to.The pressure was set to 250 mmhg.The actual pressure reading on the device was 60 mmhg.There was no patient involvement.Due diligence is complete, there is no additional information available.
 
Additional Manufacturer Narrative
An investigation into the reported event has been initiated under (b)(4).Once the investigation has been completed, a follow up report will be submitted with the investigation results and any actions taken by the manufacturer.
 
Additional Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).The following sections have been updated: b4, b5, d2, d4, g1, g3, g6, h1, h2, h3, h4, h6, and h11.Review of the most recent repair record identified no repairs related to the reported event.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
There was no additional information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
A.T.S. 4000TS TOURNIQUET W/HOSES AND LOP SENSOR (PERSONALIZED PRESSURE®)
Common Device Name
TOURNIQUET, PNEUMATIC
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key19269756
Report Number0001526350-2024-00551
Device Sequence Number7384094
Product Code KCY
UDI-Device Identifier00889024376731
UDI-Public(01)00889024376731(11)190604(10)64429472
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K123553
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 09/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number60400010100
Device Lot Number64429472
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2024
Supplement Date Received by Manufacturer09/09/2024
Initial Report FDA Received Date05/08/2024
Supplement Report FDA Received Date09/10/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/14/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-