This event has been recorded by zimmer biomet under (b)(4).The following sections have been corrected/updated: b4, b5, d2, d4, g1, g3, g6, h1, h2, h3, h4, h6, and h11.Review of the most recent repair record determined the device was not holding pressure, pinched reservoir.The reservoir was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
During the calibration process, they applied pressure to the ats, however the pressure source was fully compressed with little to no back pressure from the ats.The device passes calibration on the blue side (cuff side), however going down the red side (reservoir side), the multiple syringes that they used as a pressure source were not holding pressure on the 250 test or more.The pressure setting did not match the actual pressure it was set to.The pressure was set to 250 mmhg.The actual pressure reading on the device was 60 mmhg.There was no patient involvement.Due diligence is complete, there is no additional information available.
|