• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. APOLLORF¿ XL90, ASPIRATING ABLATOR 90°; ELECTROSURG CUT/COAG DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. APOLLORF¿ XL90, ASPIRATING ABLATOR 90°; ELECTROSURG CUT/COAG DEVICE Back to Search Results
Model Number APOLLORF¿ XL90, ASPIRATING ABLATOR 90°
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/15/2023
Type of Reportable Event Malfunction
Event or Problem Description
On 04/15/2024, it was reported by a sales representative via (b)(4) that an ar-9821 apollo is running too hot.This occurred during use in a case with no patient effect.No patient harm confirmed.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The complaint allegation is confirmed.One unpackaged ar-9811 apollo batch 74333712 was returned for investigation.Functional testing was not performed due to the damage to the device.Visual inspection noted that the tip of the device appears to have been overheated.The most likely cause is attributed to misuse due to use error caused by insufficient fluid flow or prolonged activation.Per dfu-0242-7 at revision 0.E.Precautions.5.Do not use excessive force when inserting or removing the rf probe.Do not insert the rf probe into an obstructed passageway.Patient injury and or product damage may occur.6.Do not use an rf probe as a lever to enlarge the surgical site or gain access to tissue, which may result in a bent or damaged rf probe.7.Always keep the tip of the active rf probe completely immersed in conductive irrigation fluid (saline, ringers¿ lactate, etc.), regardless of the type of arthroscopic surgery being performed.Do not use pre-warmed conductive irrigation fluids as this may result in tissue damage or thermal injury.8.A continuous flow of irrigation fluid is necessary during use of the apollo probes.Fluid flow assists in removing tissue debris and reducing the intra-joint temperature between activations.Rf probes utilized in stagnant fluid can result in heated irrigant in the joint, which can result in tissue damage inside the joint, skin burns near portals or result in damage to the probe.13.Prolonged ablation should be avoided.Intermittent pauses in ablation are recommended, to allow warm fluid and tissue debris to be evacuated from the joint.Prolonged probe activation while using the suction feature may result in elevated shaft or suction tubing temperatures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APOLLORF¿ XL90, ASPIRATING ABLATOR 90°
Common Device Name
ELECTROSURG CUT/COAG DEVICE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19270293
Report Number1220246-2024-02964
Device Sequence Number20609843
Product Code GEI
UDI-Device Identifier00888867267220
UDI-Public00888867267220
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K161581
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAPOLLORF¿ XL90, ASPIRATING ABLATOR 90°
Device Catalogue NumberAR-9821
Device Lot Number74333712
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2024
Supplement Date Received by Manufacturer08/07/2024
Initial Report FDA Received Date05/08/2024
Supplement Report FDA Received Date08/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-