| Model Number |
APOLLORF¿ XL90, ASPIRATING ABLATOR 90° |
| Medical Device Problem Code |
Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 04/15/2024, it was reported by a sales representative via (b)(4) that an ar-9821 apollo is running too hot.This occurred during use in a case with no patient effect.No patient harm confirmed.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint allegation is confirmed.One unpackaged ar-9811 apollo batch 74333712 was returned for investigation.Functional testing was not performed due to the damage to the device.Visual inspection noted that the tip of the device appears to have been overheated.The most likely cause is attributed to misuse due to use error caused by insufficient fluid flow or prolonged activation.Per dfu-0242-7 at revision 0.E.Precautions.5.Do not use excessive force when inserting or removing the rf probe.Do not insert the rf probe into an obstructed passageway.Patient injury and or product damage may occur.6.Do not use an rf probe as a lever to enlarge the surgical site or gain access to tissue, which may result in a bent or damaged rf probe.7.Always keep the tip of the active rf probe completely immersed in conductive irrigation fluid (saline, ringers¿ lactate, etc.), regardless of the type of arthroscopic surgery being performed.Do not use pre-warmed conductive irrigation fluids as this may result in tissue damage or thermal injury.8.A continuous flow of irrigation fluid is necessary during use of the apollo probes.Fluid flow assists in removing tissue debris and reducing the intra-joint temperature between activations.Rf probes utilized in stagnant fluid can result in heated irrigant in the joint, which can result in tissue damage inside the joint, skin burns near portals or result in damage to the probe.13.Prolonged ablation should be avoided.Intermittent pauses in ablation are recommended, to allow warm fluid and tissue debris to be evacuated from the joint.Prolonged probe activation while using the suction feature may result in elevated shaft or suction tubing temperatures.
|
| |
|
Search Alerts/Recalls
|