• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US MBT CEM TIBIAL TRAY SZ4; MBT TIBIAL TRAY : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US MBT CEM TIBIAL TRAY SZ4; MBT TIBIAL TRAY : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 129431140
Medical Device Problem Codes Loss of or Failure to Bond (1068); Use of Device Problem (1670)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994)
Date of Event 04/29/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Revision of lcs implants.Implants loosened due to poor positioning during initial primary procedure.There was metalosis showing in the joint.Implants were taken out very easily.Exact date unknown.
 
Event or Problem Description
Additional information was received.Indicating, that there was loosening between the implant and cement.
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Corrected: h6: (device code).Pe code, was updated from implant loosening: interface: unknown to implant loosening: interface: cement to implant.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, ¿revision of lcs implants.Original surgery year was [year].Implants loosened due to poor positioning during initial primary procedure.There was metalosis showing in the joint.Implants were taken out very easily.Exact date unknown¿ the product was not returned to depuy synthes, however photos were provided for review.See attachments "revision (b)(6).Jpg" and "revision (b)(6)2.Jpg" review of the photographic evidence shows light marks on the the surface of the rotating platform.However, due to the angle in which the photograph is taken it is not possible to observe the fixation surface of the implant.Additionally, based on the available information, the reported loose condition cannot be confirmed and it is not possible to determine the exact position in which the mbt cem tibial tray sz4 was implanted.A manufacturing record evaluation was performed for the finished device [129431140 / 3456433] number, and no non-conformances / manufacturing irregularities were identified during manufacturing.The overall complaint was not confirmed as the observed condition of the mbt cem tibial tray sz4 would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device [129431140 / 3456433] number, and no non-conformances / manufacturing irregularities were identified during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MBT CEM TIBIAL TRAY SZ4
Common Device Name
MBT TIBIAL TRAY : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19272053
Report Number1818910-2024-10183
Device Sequence Number15212349
Product Code NJL
UDI-Device Identifier10603295025610
UDI-Public10603295025610
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2017
Device Catalogue Number129431140
Device Lot Number3456433
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2024
Supplement Date Received by Manufacturer05/07/2024
05/21/2024
Initial Report FDA Received Date05/08/2024
Supplement Report FDA Received Date05/16/2024
05/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/30/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LCS COMP RP INSERT STD+ 10MM; LCS COMPLETE FEM CEM L STD+
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-