Catalog Number MCR091030
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/25/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Section d.2b: procode is krd/hcg.Section h.3: the device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.A review of manufacturing documentation associated with this lot (31177297) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.This is one of 3 products involved with the reported complaint.The associated manufacturer report numbers are: 3008114965-2024-00458, and 3008114965-2024-00459.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
The healthcare professional reported that during coil embolization of a middle meningeal artery (mma), two coils, a 1mm x 3cm cerepak freeform mini (mcr091030 / 31177297) and a 1.5mm x 2.5cm cerepak freeform mini (mcr091525 / 31149187) did not detach correctly.The cerepak tm detachment handle (mdh1 / 102109430) as well as the manual break were attempts and neither method worked.There was no report of any negative patient impact.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to include the additional event information received on 17-may-2024.[additional information]: on 17-may-2024, additional information was received.Per the information, the target vessel of the procedure was the middle meningeal artery (mma).The coils failed to detach via the detachment handle, then manual break was attempted unsuccessfully.When the coils were removed after the unsuccessful detachment attempts, they were still attached to their respective delivery systems.The coils were not stretched when they were removed.The microcatheter used with the two coils was a plato® 17 microcatheter (scientia vascular).There was no evidence of blood flow interruption as a result of the reported issue.The procedure was successfully completed; the handle was replaced and the two coils were replaced with smaller sizes.The information confirmed there was no negative patient impact and there was no procedure delay.Updated sections: b.4, g.3, g.6, h.2, h.11, and concomitant products.This is one of 3 products involved with the reported complaint.The associated manufacturer report numbers are: 3008114965-2024-00457, 3008114965-2024-00458, and 3008114965-2024-00459.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report the investigational finding of the returned device.The complaint product was returned and received for evaluation and analysis.The investigation is documented below.Investigation summary: a non-sterile 1mm x 3cm cerepak freeform mini was received contained in the decontamination pouch.Visual inspection was performed.The embolic coil was observed still attached to the delivery system.The manual break was attempted at the proper location, and the puller wire was exposed.The distal end of the delivery system was not returned for evaluation.The embolic coil was inspected under magnification, and the coil was found kinked.No damages were noted at the proximal key.No unintentional weld was noted on the loop hole.No contamination was noted at the distal end.A manufacturing record evaluation was performed for the finished device 31177297 number, and no non-conformances related to the malfunction were identified.The issue reported was confirmed based on the detachment attempts, the broken condition of the manual break, and the embolic coil being still attached.With the limited information available and the evidence obtained from the device inspection, there is no clear insight into the root cause and/or exact contributing factors that may have resulted in the observed failure mode.It should be noted that product failure is multifactorial.A capa is currently investigating failure to detach.The kinked conditions of the embolic coil were not originally reported; the exact time of occurrence cannot be determined; therefore, they are not considered related to the issue reported.It should be noted that product failure is multifactorial.The instructions for use (ifu) contains the following recommendations: remove the detacher from the protective packaging and place it within the sterile field.The detacher is packaged separately for single patient use only.Confirm again under fluoroscopy that the coil alignment marker of the delivery tube creates a ¿t¿ with the proximal marker of the microcatheter.Verify that the rhv is firmly locked around the delivery tube to ensure that the coil does not move during connection process.Ensure the delivery tube is straight between the rhv and the detacher.Hold the delivery tube approximately 3 inches from proximal end and bring the detacher over the delivery tube until a firm stop is felt.Maintain the delivery tube position within the detacher and use a thumb to pull back the slider on the detacher.Continue to pull on the slider until end of travel or a click is heard.Gently release the slider until it returns to starting position or a second click is heard.Remove detacher by slowly withdrawing it from the delivery tube.Note: detacher must remain in sterile field until the procedure is completed.Fluoroscopically verify that the coil is detached by gently pulling back the delivery tube approximately 1cm.If the coil has detached, remove the delivery tube from the microcatheter.If the coil has not detached, repeat steps 3-7.If 2 unsuccessful detachments occur, discard the detacher and proceed to the manual break section.Repeat steps 1-8 until the procedure is completed.Discard the detacher.Caution: always perform fluoroscopic verification of detachment prior to withdrawing the delivery system fully.Failure to do so may lead to an embolic complication.After detaching the coil, remove the delivery tube from the microcatheter and discard.Repeat the above sequence for all additional coils until the procedure is complete.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Updated sections: b.4, g.3, g.6.H.2, h.3, h.6, and h.11.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that the product analysis lab received the complaint device on 30-may-2024.A supplemental 3500a report will be submitted once the product investigation has been completed.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREEFORM MINI 1MM X 3CM
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS, INC.
calle circuito int norte 1820
parque industrial salvacar
juarez 32574
MX   32574
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
303552-6892
MDR Report Key19272090
Report Number3008114965-2024-00457
Device Sequence Number12504624
Product Code KRD
UDI-Device Identifier10886704082934
UDI-Public10886704082934
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K220040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Catalogue NumberMCR091030
Device Lot Number31177297
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2024
Supplement Date Received by Manufacturer05/17/2024
05/30/2024
06/21/2024
Initial Report FDA Received Date05/08/2024
Supplement Report FDA Received Date05/17/2024
05/30/2024
06/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MECHANICAL DETACHMENT HANDLE; PLATO® 17 MICROCATHETER (SCIENTIA VASCULAR)