| Model Number |
324910 |
| Medical Device Problem Code |
Volume Accuracy Problem (1675)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
3 syringes were affected by this issue.
|
| |
|
Event or Problem Description
|
|
Consumer reported, the first line on the barrel is different between 3 syringes.Stated, he did use the syringes.Lot: 3268200 catalog: 324910 date of event: unknown samples: yes cl.
|
| |
|
Additional Manufacturer Narrative
|
|
No samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st.Complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformance's were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
|
| |
|