This device is classified as import for export, therefore 510k is not applicable.Model fnl-10rp3 is available in the usa with a 510k number k951196.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ocular cover glass cloudy (not clear view).Based on the result, we concluded that it was caused due to the moisture condensation in the ocular cover glass.In addition, our technician confirmed that the u/d knob broken, the insertion flexible tube perforated, the insertion flexible tube buckled, and the insertion flexible tube crushed; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|