| Model Number |
45 DEG CURVE LEFT, QUICKPASS LASSO |
| Medical Device Problem Code |
Material Frayed (1262)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, it was reported by a sales representative via sems (b)(4) that an ar-6068-45l quickpass lassos wire frayed.It essentially split into two.This occurred during a labral repair.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause is attributed to misuse due to prying/leveraging the device during use.
|
| |
|
Search Alerts/Recalls
|