| Model Number |
DRIVER SHAFT, T6, NON-RETAINING, AO |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 04/16/2024, it was reported by a sales representative via (b)(4) that (2) ar-18800-03 driver shafts and (2) ar-18800-04 driver shafts were warped and breaking.This occurred during an orif ankle procedure on (b)(6)2024.No additional information was provided, and additional information has been asked.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 complaint allegation is confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause is attributed to user error of the device due to over-torquing/over-engaging the driver within the screw head.
|
| |
|