• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC 10 SHOOTER SAEED MULTI-BAND LIGATOR; MND, LIGATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC 10 SHOOTER SAEED MULTI-BAND LIGATOR; MND, LIGATOR Back to Search Results
Catalog Number MBL-U-10
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.A review of the device history record could not be conducted because the lot number was not provided.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Band deployment difficulty can occur if the endoscope accessory channel is compromised.In these cases, the endoscope accessory channel collapses, restricting the trigger cord and preventing proper band deployment.The instructions for use contain the following statement: "use of an endoscope in a sound state of repair is a prerequisite for a successful multi-band ligation procedure." band deployment difficulty can also occur if the trigger cord is not properly seated in the handle assembly.The instructions for use advise the user: "note: knot must be seated into hole or handle will not function properly." a precaution in the instructions for use state: "it is vital that the integrity of the working channel is intact as grooves or other obstructions in the working channel can potentially cause the string to catch, resulting in band deployment difficulty." it is possible that this could lead to band deployment difficulty and damage to the trigger cord as well.Prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an internal hemorrhoid ligation, the physician used a cook 10 shooter saeed multi-band ligator.It was reported that the user detected the trigger cord entangled and stuck during the procedure which caused the bands to not be deployed.User re-installed the device and completed the procedure.Other than the deployed bands, a section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Band deployment difficulty can occur if the endoscope accessory channel is compromised.In these cases, the endoscope accessory channel collapses, restricting the trigger cord and preventing proper band deployment.The instructions for use contain the following statement: "use of an endoscope in a sound state of repair is a prerequisite for a successful multi-band ligation procedure." band deployment difficulty can also occur if the trigger cord is not properly seated in the handle assembly.The instructions for use advise the user: "note: knot must be seated into hole or handle will not function properly." a precaution in the instructions for use state: "it is vital that the integrity of the working channel is intact as grooves or other obstructions in the working channel can potentially cause the string to catch, resulting in band deployment difficulty." it is possible that this could lead to band deployment difficulty and damage to the trigger cord as well.Prior to distribution, all 10 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10 SHOOTER SAEED MULTI-BAND LIGATOR
Common Device Name
MND, LIGATOR
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key19278572
Report Number1037905-2024-00247
Device Sequence Number15037548
Product Code MND
UDI-Device Identifier00827002319196
UDI-Public(01)00827002319196(17)240928(10)W4769258
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/28/2024
Device Catalogue NumberMBL-U-10
Device Lot NumberW4769258
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/16/2024
Supplement Date Received by Manufacturer05/10/2024
Initial Report FDA Received Date05/09/2024
Supplement Report FDA Received Date06/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCOPE, UNKNOWN MAKE AND MODEL.
Patient Age23 YR
Patient SexFemale
-
-