Patient #(b)(6) : 93-year-old male was implanted with lcp broad, curved plates 4.5/5.0 (226.682) in femur with dps screws (unknown quantity) with other adjunctive fixation devices (2 cables (synthes), fibular strut).Intra-operative complication: fracture comminution experienced intra-op related to synthes procedure.Treatment: surgical treatment for hardware removal performed for removal of all hardware 12 weeks post-op.Outcome: unknown.This report is for a 4.5mm curved broad lcp® plate 18 holes/336mm.This is report 1 of 5 for (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D2: additional procode: hwc d9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a synthes employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.H6 component code: g07002 ¿ device not returned.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|