• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSCOR.INC ANGIODYNAMICS SAFESHEATH® ULTRA LITE; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSCOR.INC ANGIODYNAMICS SAFESHEATH® ULTRA LITE; INTRODUCER, CATHETER Back to Search Results
Model Number VI7
Medical Device Problem Code Material Too Rigid or Stiff (1544)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Conclusion not yet available, evaluation in process.A follow-up will be submitted as soon as the investigation is complete.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Event or Problem Description
The clear plastic ring on the peel away sheath are not breaking free with the sheath.In two cases the ring separated in pieces and broke off in the sheath.In the third case, it didn't break open like it should.The procedures were completed with these devices and the patients did not experience any adverse effects, harm, or require medical intervention as a result of these incidents.
 
Additional Manufacturer Narrative
The following sections were updated in follow-up 1.No product was provided for analysis.Procedure completed successfully with same device.There was no adverse event with the patient reported.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspections.No product was returned to oscor inc.For evaluation; however, a complaint notification was provided.Without the return of the actual device in question for evaluation, oscor inc.Is unable to determine the exact cause for this incident.At this time, it is not possible to assign a definitive root cause for the event as reported.The complaint is non-verifiable as the product was not returned for evaluation.No corrections or corrective actions will be taken.Device was not returned for analysis and review of the manufacturing documents did not reveal any discrepancies that would have contributed to this event.This complaint is non-verifiable.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYNAMICS SAFESHEATH® ULTRA LITE
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
OSCOR.INC
3816 desoto blvd.
palm harbor FL 34683
Manufacturer (Section G)
OSCOR.INC
3816 desoto blvd.
palm harbor FL 34683
Manufacturer Contact
daniel naut
3816 desoto blvd.
palm harbor, FL 34683
7279372511
MDR Report Key19281687
Report Number1035166-2024-00052
Device Sequence Number18040457
Product Code DYB
UDI-Device Identifier25051684026267
UDI-Public25051684026267
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K073100
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVI7
Device Catalogue NumberVI7
Device Lot NumberDP-19019
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/10/2024
Supplement Date Received by Manufacturer06/18/2024
Initial Report FDA Received Date05/09/2024
Supplement Report FDA Received Date06/18/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-