D10: concomitant devices: 1304221, 164-02-11 - element-stem, collarless w/ha, high offset, sz 11, 1704809, 120-65-15 - bone screw 6.5mm dia x 15mm long, 1738909, 120-65-30 - bone screw 6.5mm dia x 30mm long, 1987951, 142-32-93 - cocr fem head 32mm -3.5 offset 12/14, 2001188, 180-01-52 - nv crown cup clstr hole 52mm group 2.The device was not returned for evaluation and no photographs or x-rays were provided for review; therefore the reported event cannot be confirmed through analysis.No operative notes were provided; therefore, a review of device usage and technique could not be performed.No information concerning patient conditions, co-morbidities, or other health or anatomical concerns was provided and it is therefore unknown if patient-related factors may have caused or contributed to the reported event.The cause of the reported event cannot be determined based on the information made available.Should additional, relevant information become available, a supplemental report will be filed accordingly.
|
|
H3: based on the available information, the patient involved meets the following risk criteria for early prosthesis wear and/or osteolysis as specified in the hhe: implanted with a lateralized liner and combination of largest available femoral head and/or thinnest available acetabular liner was used.The most likely cause for the reported prosthesis wear is a combination of the risk factors specified in the hhe.However, this cannot be confirmed as the devices were not available for evaluation, and images and radiographs were not provided.H6: corrected medical device clinical code.
|