MEDTRONIC IRELAND ATTAIN SELECT II; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 6248V-130
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Dyspnea (1816); Pericardial Effusion (3271)
Date of Event 04/17/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during an implant procedure, the coronary sinus was accessed with two delivery catheters and there was an attempt to place a left ventricular (lv) lead in several different branches with no success.It was noted that there was intermittent capture at high outputs.It was decided to do an occluded venogram fluoroscopy save.The lv lead was placed in two other branches as well and there was no capture.The decision was made to abort the case as the patient¿s blood pressure was getting low and there were no branches that were usable.The two delivery catheters and lv lead were removed from the patient.The patient then complained of difficulty breathing.The physician called for an ultrasound machine and then a code.The patient experienced a pericardial effusion and a pericardiocentesis procedure was performed with no complications.It was noted that the existing leads were never disconnected, and the existing implantable pulse generator (ipg) pacing mode was changed.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN SELECT II
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key19282811
Report Number9612164-2024-02230
Device Sequence Number7430600
Product Code DQY
UDI-Device Identifier00763000160517
UDI-Public00763000160517
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K123153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/16/2024
Device Model Number6248V-130
Device Catalogue Number6248V-130
Device Lot Number0011271081
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2024
Initial Report FDA Received Date05/09/2024
Date Device Manufactured06/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient Age84 YR
Patient SexFemale
-
-