| Model Number |
ASP-0001 |
| Medical Device Problem Codes |
Material Frayed (1262); Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
Capsular Bag Tear (2639)
|
| Date of Event |
04/25/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The doctor experienced capsule tears using allegro sp ia handpieces.Four separate patients were impacted by 4 different devices.All patients recovered with no intervention required.The procedure was completed as planned.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is 4 of 4.The dfu for the device instructs the user to inspect the device before use and if damage is present not to use the device and contact the manufactuer.The doctor chose to use the device even though it appeared to be out of specification.
|
| |
|
Search Alerts/Recalls
|