EXACTECH, INC. CC TIBIAL INSERT SZ 2, 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
208-22-15 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Emotional Changes (1831); Failure of Implant (1924); Joint Laxity (4526); Implant Pain (4561)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: concomitant devices: (b)(6), 204-62-05 - tibial augment block 1/2-sz 2 5mm.(b)(6), 204-62-05 - tibial augment block 1/2-sz 2 5mm.(b)(6), 204-04-22 - trapezoid tibial tray sz 1f/2t, 2f/2t.(b)(6), 208-01-02 - cc femoral sz 2.(b)(6), 200-02-35 - three peg patella 35mm.(b)(6), 208-09-05 - cc stem ext adaptor 5 degree.(b)(6), 204-34-04 - fluted stem extension 40l x 14 mm.(b)(6), 204-34-04 - fluted stem extension 40l x 14 mm.The device was not returned for evaluation and no photographs or x-rays were provided for review; therefore the reported event cannot be confirmed through analysis.No operative notes were provided; therefore, a review of device usage and technique could not be performed.No information concerning patient conditions, co-morbidities, or other health or anatomical concerns was provided and it is therefore unknown if patient-related factors may have caused or contributed to the reported event.The cause of the reported event cannot be determined based on the information made available.Should additional, relevant information become available, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that approximately 101 months after a right total knee replacement procedure, the patient may require a future revision procedure to address polyethylene wear.The patient has reported mental anguish and agony, pain, and instability.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: corrected the following: health effect - clinical code, health effect - impact code, type of investigation manufacturer narrative updated: the reason for the reported event in case-(b)(4) cannot be confirmed from the information provided but may be due to prosthesis wear and/or inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.
|
| |
|
Search Alerts/Recalls
|
|
|