MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC DUODERM; DRESSING,WOUND,OCCLUSIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC DUODERM; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Model Number 187900
Medical Device Problem Codes Delivered as Unsterile Product (1421); Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The nurse reported that foreign particles were found on the dressings.The product was not used by the patient.A photograph depicting the issue was received from the complainant.
 
Additional Manufacturer Narrative
Device 2 of 3.E1: complainant city: (b)(6).Complainant state: (b)(6).Complainant country: peru.Name of affiliation: cardio perfusion e.I.R.L patient city: (b)(6).Patient state: (b)(6).Patient country: peru.Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number, reporting site: 1049092, manufacturing site: 9618003.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Additional Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions photograph, video and/or physical sample evaluation: there is a photograph associated with this case and in this, the reported defect can be seen.No unused return sample was expected.Batch record revision results: lot 3d04593 was manufactured on 08/ may 2023, in doyen b manufacturing line, with a total of (b)(4) market units (mkus).The complaints investigator performed a batch record review on 05/10/2024, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, system application product (sap) material identification (id) 1704764 and manufacturing order (b)(4).Review of the batch record showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancy related to this issue were found within the documentation.Historical complaints review: on 05/10/2024, complaints investigator ran a query in database in order to verify the complaints reported for the 3d04593 lot for the malfunction code ¿foreign matter within product, sterile products (e.G.Metal contamination in dressings, particulates within fluids, gels etc)¿ defect and no additional complaints were identified.Historical nonconformance review: on 05/10/2024, complaints investigator ran a query in database looking for any in process nonconformance / corrective action / preventive actions (capa) (s) associated to the malfunction code ¿foreign matter within product, sterile products (e.G.Metal contamination in dressings, particulates within fluids, gels etc)¿ defect for the lot number 3d04593 and as result, no nonconformance / corrective action / preventive actions (capa) (s) for this malfunction code were generated during the manufacturing process of the referenced lot.Defect rate analysis: there have only been 3 defective parts confirmed to date from a lot size 198,000 products.This represents a defect rate of only 0.0015% which is well within an appropriate acceptable quality level (aql) for this defect which should be 1.0 based on our test methods (tm).In addition, all the in-process testing on this lot did not find a single defective unit, which confirms that the lot is unlikely to breach an acceptable quality level (aql) of 1.0 this issue certainly appears to be an isolated incident, but more importantly to date, it is well within our accepted acceptable quality level (aql) level for this type of failure mode or defect.Conclusions: as photographs was available for this complaint issue, it was evaluated and as a result, the reported malfunction for the observed product was confirmed, however, after performing the defect rate analysis for this issue, it was observed that the quantity of reportedly affected products is within the appropriate acceptable quality level (aql) for this defect.A batch record review was completed and showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancy related to this issue were found within the documentation.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUODERM
Common Device Name
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR  91000
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR   91000
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key19286177
Report Number9618003-2024-01902
Device Sequence Number12868358
Product Code NAD
UDI-Device Identifier00768455175048
UDI-Public00768455175048
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number187900
Device Lot Number3D04593
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2024
Supplement Date Received by Manufacturer05/13/2024
Initial Report FDA Received Date05/10/2024
Supplement Report FDA Received Date06/07/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/08/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-