| Lot Number |
J001778 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by (b)(6), an onkos sales representative, that a 4.3 x 200mm drill bit fractured intraoperatively in the process of implanting a my3d acetabular reconstruction implant (case # (b)(4)).The drill bit fragment(s) reportedly remain implanted in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The incident remains under investigation.Additional findings will be submitted upon discovery via supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported by s.Steinert, an onkos sales representative, that a 4.3 x 200mm calibrated drill bit fractured intraoperatively on (b)(6) 2024 while the surgeon, doctor morgan, was drilling screw holes to implant a my3d pelvic reconstruction system (case # (b)(6)) into a female patient.The fractured drill bit fragments reportedly remain implanted in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on review of device history records, the investigation concluded that the root cause of the reported mechanical failure (fracture) of the surgical instrument was not related to the design, manufacture, and/or sterilization of the 4.3 x 200mm calibrated drill bits.Results obtained through a physical evaluation of samples from the same lot which were held at onkos indicated that the most probable root cause which led to the fractured drill bits is unintended use error involving excessive bending of the drill bits.
|
| |
|
Search Alerts/Recalls
|