| Model Number |
VICM5 13.2 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Toxic Anterior Segment Syndrome (TASS) (4469)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm vicm5 -6.00d implantable collamer lens into the patient's eye on an unknown date.Information about concomitant products used including ovd and delivery system were not provided.Toxic anterior segment syndrome (tass) was diagnosed.Per the last report dated apr-17-2024, no information is provided that there was any medical or surgical intervention performed including lens removal.No loss of visual acuity was reported.Additional information has been requested but none has been forthcoming.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Manufacturer narrative: a review of the device labeling was completed.Inflammation is identified in the labeling as a known potential adverse event following icl implantation.The dfu provides the surgeon instruction for complete ovd removal.Precaution: (6) after inserting this product, aspirate the viscoelastic substance completely from inside the eye.Do not use highly viscous viscoelastic substances that are difficult to aspirate completely.Toxic anterior segment syndrome (tass) is an acute sterile postoperative inflammation that can occur after uncomplicated or complicated intraocular surgery.While improper sterilization and contamination of surgical instruments remains the most common risk factor associated with tass, ocular viscoelastic, and improper preparation of antibiotics for intracameral injection may also lead to tass.An exact cause could not be determined as the event could be multifactorial in nature including patient and/or procedure related factors.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation 3331 - device history record (dhr) review - the device history records indicates that the product has been manufactured within the established process parameters and that there is no indication that the manufacturing and/or processing of the device contributed to the complaint issue.Corrective and/or preventive actions are being addressed through capa-24-001 and this issue will continue to be tracked and trended as part of the normal device safety committee meetings and capa monitoring.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5- additional information was later reported that the patient recovered on day eight post-op.Claim# (b)(4).
|
| |
|