| Model Number |
NM-F4432 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Unspecified Infection (1930)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The healthcare professional reports that nm-f4432 lot#0023605 implant was inserted.The implantation and removal date has not been provided.Observed during the event: mobility/infection.The device was forwarded to the manufacturer.No further patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The dhr was reviewed for lot#0023605 and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.To date, items of the same lot have been sold without any similar reported event during the period of review.The investigation for this event has not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Our manufacturing system assures, that production and process controls are in place to ensure that batches conform to the applicable specifications before they are distributed.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only.
|
| |
|