EXACTECH, INC. TRULIANT PS CEM FEM PS CEM LEFT SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-020-11-0230 |
| Medical Device Problem Codes |
Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Muscle Weakness/Atrophy (1967); Pain (1994); Discomfort (2330); Joint Laxity (4526); Swelling/ Edema (4577); Aseptic loosening (4883)
|
| Date of Event |
04/23/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending investigation.D10: (b)(6) 02-022-35-3009 - truliant tib imp ps insert sz 3 9mm (b)(6) 02-022-45-3020 - truliant tib fit tray cem sz 3f / 2t (b)(6) 200-07-32 - advanced patella 32mm 3 peg implant.
|
| |
|
Event or Problem Description
|
|
As reported, the 74 year old female patient had an initial left tka on (b)(6) 2018.The patient complained of pain and was revised on (b)(6) 2024 due to a loose femur and bad poly.There was no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b5, h6 component code, type of investigation, investigation findings, and investigation conclusions.The following sections were corrected: b2, h6 clinical code and problem code.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 76 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, loosening, pain, stiffness, swelling, discomfort, weakness and instability.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for the revision reported was likely the result of pain, which was caused by prosthesis wear, instability, and an insufficient bond between the femoral component and the bone, which led to aseptic (non-infected) femoral loosening.The reason for the reported loosening cannot be confirmed as the devices were not returned for evaluation.Failure of the cement used to secure the femoral component to the femoral bone, may have lead to loosening at the cement-bone interface and/or cement-implant interface; however, this cannot be confirmed.Prosthesis wear of the tibial insert could not be confirmed from the images provided.Possible causes for polyethylene wear for the patella include malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability or any combination of these possibilities.H6: corrected health effect and investigation clinical codes.
|
| |
|