| Model Number |
APS II SHVR HANDP,FT-SWCH CNTRL |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Event description: it was reported that the device has speed fluctuations during use and gets hot.The reported event was confirmed.The service evaluation revealed a short circuit in the motor along with damages at the suction valve.Furthermore, the locking pin is sticking, the cable is worn and the soak cap is corroded.
|
| |
|
Event or Problem Description
|
|
It was reported that the device has speed fluctuations during use and gets hot.There was no harm or adverse event for patient, operator or third party reported.No further information received.
|
| |
|