• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDO ILS 21MM, CURVED; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDO ILS 21MM, CURVED; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ECS21A
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent 5/13/2024.D4 batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a distal gastrectomy surgery, after fired, noted the staples malformed.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 6/22/2024.D4: batch #495c88.Investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the ecs21a device arrived with no apparent damage.The breakaway washer was present, cut and there were no staples present, indicating that the device achieved a full firing stroke.The device was reloaded with staples, a new washer was placed on the device and it was tested for functionality.It fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.No malformed staples were found at any time during the testing.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: the firing stroke must be completed.Do not partially fire the instrument.Incomplete firing can result in malformed staples, incomplete cut line, bleeding, and leakage from the staple line and/or difficulty removing the device.Ensure that the firing trigger is fully squeezed to ensure proper staple formation and cutting of tissue.The event described could not be confirmed as the device performed without any difficulties.A manufacturing record evaluation was performed for the finished device batch number 495c88, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO ILS 21MM, CURVED
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key19297705
Report Number3005075853-2024-03555
Device Sequence Number20541517
Product Code GDW
UDI-Device Identifier10705036003489
UDI-Public10705036003489
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K940967
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberECS21A
Device Lot NumberA9DC97
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/05/2024
Supplement Date Received by Manufacturer05/29/2024
Initial Report FDA Received Date05/13/2024
Supplement Report FDA Received Date06/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Patient SexUnknown
-
-