• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION EVOLUT PRO TRANSCATHETER AORTIC VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION EVOLUT PRO TRANSCATHETER AORTIC VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number EVOLUTPRO-23-US
Medical Device Problem Codes Calcified (1077); Gradient Increase (1270); Inadequacy of Device Shape and/or Size (1583); Patient Device Interaction Problem (4001); Thickening of Material (4056); Device Stenosis (4066)
Health Effect - Clinical Codes Aortic Valve Stenosis (1717); Thrombosis/Thrombus (4440)
Date of Event 05/06/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic has requested additional information pertaining to this reportable event.If additional reportable information is received, a supplemental report will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information that five years and nine months following the implant of this transcatheter bioprosthetic valve, valve stenosis was observed and patient-prosthesis mismatch was identified.Thrombus was reported.Root enlargement surgery was performed and the valve was explanted and replaced.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information that five years and nine months following the implant of this transcatheter bioprosthetic valve, valve stenosis was observed and patient-prosthesis mismatch was identified.Thrombus was reported.Root enlargement surgery was performed and the valve was explanted and replaced.No additional adverse patient effects were reported.Additional information was received that the gradient was 37 mmhg when the thrombus was detected.Mild hyperattenuated leaflet thickening and annular calcium were noted.The patient had pre-existing thrombocytopenia, and was on an anticoagulant medication since implant due to atrial fibrillation.No change in the patient's anticoagulant medication was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT PRO TRANSCATHETER AORTIC VALVE
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key19298549
Report Number2025587-2024-02902
Device Sequence Number6980552
Product Code NPT
UDI-Device Identifier00643169954434
UDI-Public00643169954434
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P130021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/18/2019
Device Model NumberEVOLUTPRO-23-US
Device Catalogue NumberEVOLUTPRO-23-US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2024
Supplement Date Received by Manufacturer05/23/2024
Initial Report FDA Received Date05/13/2024
Supplement Report FDA Received Date06/05/2024
Date Device Manufactured06/05/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age70 YR
Patient SexFemale
-
-