ARTHREX, INC. ACL TIGHTROPE; SMOOTH FIXATION PIN Back to Search Results
Model Number ACL TIGHTROPE
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on (b)(6) 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The device was not returned for evaluation and therefore the alleged complaint event could not be confirmed.The most likely cause for the reported failure can be attributed to misuse/mishandling due to damage to the device.However, due to the device not being returned, we are unable to confirm the reported event.
 
Event or Problem Description
It was reported that during a knee surgery the device broke off inside the patient.The broken pieces were retrieved from patient.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACL TIGHTROPE
Common Device Name
SMOOTH FIXATION PIN
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19299379
Report Number1220246-2024-03195
Device Sequence Number6883323
Product Code MBI
UDI-Device Identifier00888867019614
UDI-Public00888867019614
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K231857
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberACL TIGHTROPE
Device Catalogue NumberAR-1588T
Device Lot Number15019657
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/06/2023
Initial Report FDA Received Date05/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-