• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. KNEE IMPLANT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. KNEE IMPLANT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number PS TIBIAL INSERTS SZ 2, 9MM
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approximately 189 months after a left total knee replacement procedure, the patient may require revision surgery for prosthesis wear.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
Concomitant devices: unknown the device was not returned for evaluation and no photographs or x-rays were provided for review; therefore the reported event cannot be confirmed through analysis.No operative notes were provided; therefore, a review of device usage and technique could not be performed.No information concerning patient conditions, co-morbidities, or other health or anatomical concerns was provided and it is therefore unknown if patient-related factors may have caused or contributed to the reported event.The cause of the reported event cannot be determined based on the information made available.Should additional, relevant information become available, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
The reason for the reported event cannot be confirmed from the information provided but may be due to prosthesis wear and/or inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KNEE IMPLANT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key19302785
Report Number1038671-2024-01214
Device Sequence Number6980678
Product Code JWH
UDI-Device Identifier10885862048226
UDI-Public10885862048226
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/20/2016
Device Model NumberPS TIBIAL INSERTS SZ 2, 9MM
Device Catalogue Number204-22-09
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2024
Supplement Date Received by Manufacturer03/05/2025
Initial Report FDA Received Date05/13/2024
Supplement Report FDA Received Date03/05/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-