| Model Number |
CM-99145BRN |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned.Production records were reviewed.There were no nonconformities noted with the lot during production.There were no nonconformities with the raw material used.All finished goods testing requirements were met prior to release.There was no evidence that the device failed to meet specifications and the report could not substantiated.There is a possibility of using the incorrect surgical method for implanting the inserter.Furthermore, if the drill channel is too small or the bone quality is not of acceptable form, this kind of failure is possible.A cause for the event cannot be established.This report and use of categorical definitions required by fda 3500a does not constitute an admission by riverpoint medical or its employees that riverpoint medical or its employees has cause or contributed to the event described in the report.Riverpoint medical filed this information to comply with the medical device reporting regulation 21 cfr 803.If additional information is provided to riverpoint medical regarding this event, a supplementary 3500a form will be submitted as required by fda.
|
| |
|
Event or Problem Description
|
|
According to the reporter, "it was reported that during the surgery, the inserter of this complaint product couldn't be pulled out after the surgeon inserted the anchor into the patient's burr hole.And the surgeon pulled the inserter to pull it out from the patient's burr hole, then tip of the inserter was fractured and remaining in the patient's burr hole.Fortunately, the fractured piece of inserter could be removed, so there is no remains in the patient's body.".
|
| |
|
Search Alerts/Recalls
|