| Model Number |
P200DNUS |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3.Date of event: unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the product had a problem with the battery level lamp.Event occurred before patient use, during testing.There was no patient involvement and no harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received.No abnormality related to the reported event was confirmed on the product's appearance.It was confirmed that the low battery voltage alarm was flashing and sounding even after replacing the battery with a new one.The cause is a failure of the main alarm board.The product was returned to the customer without being repaired at the customer's request.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
|
| |
|