• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL LTD. PNEUPAC PARAPAC VENTILATOR W/ ALARMS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL LTD. PNEUPAC PARAPAC VENTILATOR W/ ALARMS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Model Number P200DNUS
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3.Date of event: unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that the product had a problem with the battery level lamp.Event occurred before patient use, during testing.There was no patient involvement and no harm/adverse event reported.
 
Additional Manufacturer Narrative
One device was received.No abnormality related to the reported event was confirmed on the product's appearance.It was confirmed that the low battery voltage alarm was flashing and sounding even after replacing the battery with a new one.The cause is a failure of the main alarm board.The product was returned to the customer without being repaired at the customer's request.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC PARAPAC VENTILATOR W/ ALARMS
Common Device Name
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL LTD.
enterprise way
luton LU3 4 BU
UK  LU3 4BU
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key19305451
Report Number9611178-2024-00082
Device Sequence Number7321239
Product Code BTL
UDI-Device Identifier35019315072738
UDI-Public(01)35019315072738(11)090612
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K020899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberP200DNUS
Device Catalogue NumberP200D/NJJP
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/15/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2024
Supplement Date Received by Manufacturer05/19/2024
Initial Report FDA Received Date05/13/2024
Supplement Report FDA Received Date06/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/12/2009
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-