Catalog Number 3543007MC
Medical Device Problem Code Gel Leak (1267)
Health Effect - Clinical Code Granuloma (1876)
Date of Event 01/01/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent a breast augmentation surgery with implantation of a 300cc mentor memorygel breast implant prosthesis.Post-operatively, the patient experienced bilaterally enlarged lymph nodes with pain, throbbing, and discomfort.Biopsies were performed and it was determined that the patient developed silicone lymphadenopathy in both armpits.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.The date of implantation and event were provided to mentor as a best estimate.Mentor became aware that a discrepancy was observed when the information reported included a date of implantation that occurred prior to the manufacturing date of the breast implant.As implant date as provided as an approximate date and they do not know what the exact date was, for further follow-ups are necessary.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.  as such, the investigation will be closed.  if the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.  this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.  this report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.  if certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: biospy manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key19305722
Report Number1645337-2024-05701
Device Sequence Number7351777
Product Code FTR
UDI-Device Identifier00081317004394
UDI-Public00081317004394
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2019
Device Catalogue Number3543007MC
Device Lot Number6798977
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/26/2024
Initial Report FDA Received Date05/13/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/03/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-