• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN AG JUAREZ (MMJ) ESOFLIP; ESOPHAGEAL DILATOR BALLOON WITH OR WITHOUT ELE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN AG JUAREZ (MMJ) ESOFLIP; ESOPHAGEAL DILATOR BALLOON WITH OR WITHOUT ELE Back to Search Results
Model Number ES-330
Medical Device Problem Codes Inflation Problem (1310); Leak/Splash (1354); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation summary: functional testing of the return device confirmed a manufacturing fault.The root cause of the observed condition was determined to be a result of a manufacturing activity.A process improvement has been initiated to prevent this condition from recurring.A review of the device history record indicates the product was released meeting all medtronic quality release specifications at the time of manufacture.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, all 4 catheters did not inflate, and did not display data on the screen and some were leaking.The customer pulled another unit from a different box and continued without issue.There was no patient or user harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESOFLIP
Common Device Name
ESOPHAGEAL DILATOR BALLOON WITH OR WITHOUT ELE
Manufacturer (Section D)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX  32575
Manufacturer (Section G)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX   32575
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19305784
Report Number9681384-2024-00082
Device Sequence Number17975078
Product Code PID
UDI-Device Identifier10884521809451
UDI-Public10884521809451
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K132337
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 05/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/22/2025
Device Model NumberES-330
Device Catalogue NumberES-330
Device Lot Number23H0624JZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2024
Initial Report FDA Received Date05/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3004904811-04-29-2024-00
Patient Sequence Number1
Patient SexUnknown
-
-