• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICM5 12.6
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Toxic Anterior Segment Syndrome (TASS) (4469)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6 - type of investigation - lens work order search: one similar complaint type event reported for units within the same lot.Manufacturer narrative: a review of the device labeling was completed.Inflammation is identified in the labeling as a known potential adverse event following icl implantation.The dfu provides the surgeon instruction for complete ovd removal.Precaution: (6) after inserting this product, aspirate the viscoelastic substance completely from inside the eye.Do not use highly viscous viscoelastic substances that are difficult to aspirate completely.Toxic anterior segment syndrome (tass) is an acute sterile postoperative inflammation that can occur after uncomplicated or complicated intraocular surgery.While improper sterilization and contamination of surgical instruments remains the most common risk factor associated with tass, ocular viscoelastic, and improper preparation of antibiotics for intracameral injection may also lead to tass.An exact cause could not be determined as the event could be multifactorial in nature including patient and/or procedure related factors.Claim# (b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 12.6mm vicm5 -6.50d implantable collamer lens into the patient's eye.Information about concomitant products used including ovd and delivery system were not provided.Toxic anterior segment syndrome (tass) was diagnosed.Per the last report dated (b)(6) 2024, no information is provided that there was any medical or surgical intervention performed.No loss of visual acuity was reported.Additional information has been requested but none has been forthcoming.If additional information is received a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
B5- the reporter indicated that the surgeon implanted a 12.6mm vicm5 -6.50d implantable collamer lens into the patient's right eye (od) on (b)(6) 2024.Concomitant products used intraoperatively include "purified sodium hyaluronate" ovd and the msi injector system.Toxic anterior segment syndrome (tass) was diagnosed.Per the last report dated (b)(6) 2024, the status of the eye was reported as "under observation", however no information was provided that there was any medical or surgical intervention performed.No loss of visual acuity was reported.Claim # (b)(4).
 
Additional Manufacturer Narrative
H6 device history record (dhr) review: based on the results of the investigation, the dhr review indicates that the product has been manufactured within the established process parameters and there is no indication that the manufacturing and/or processing of the device contributed to the complaint issue.(b)(4).
 
Additional Manufacturer Narrative
B5 - the reporter indicated that the surgeon implanted a 12.6mm vicm5 -6.50d implantable collamer lens into the patient's right eye (od) on (b)(6) 2024.It was reported that at 29 days post-implantation, inflammatory recovery was observed.Concomitant products used intraoperatively include "purified sodium hyaluronate" ovd and the msi injector system.Toxic anterior segment syndrome (tass) was diagnosed.Per the last report dated (b)(6) 2024, the status of the eye was reported as "under observation", however no information was provided that there was any medical or surgical intervention performed.No loss of visual acuity was reported.Claim # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key19305852
Report Number2023826-2024-02129
Device Sequence Number20480490
Product Code MTA
UDI-Device Identifier00840311300532
UDI-Public00840311300532
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Model NumberVICM5 12.6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2024
Supplement Date Received by Manufacturer09/22/2024
05/01/2024
09/12/2025
Initial Report FDA Received Date05/13/2024
Supplement Report FDA Received Date09/25/2024
09/25/2024
10/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-