• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ZT TRIAL SLEEVE 18F SML; FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ZT TRIAL SLEEVE 18F SML; FEMORAL TRIALS Back to Search Results
Catalog Number 257618120
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Integrity Problem (2978)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that srom trial sleeve would not fully seat in inserter.It appears the trial was damaged.This did not cause a delay in the case and the product was not broken into pieces.
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZT TRIAL SLEEVE 18F SML
Common Device Name
FEMORAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19308730
Report Number1818910-2024-10516
Device Sequence Number15100678
Product Code HWT
UDI-Device Identifier10603295146629
UDI-Public10603295146629
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number257618120
Device Lot NumberMT1007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2024
Supplement Date Received by Manufacturer05/29/2024
Initial Report FDA Received Date05/14/2024
Supplement Report FDA Received Date05/30/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/15/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
-
-